A simple graphical method to evaluate deadtime and countrate losses
Creators
- 1. Oesterreichische Akademie der Wissenschaften, Vienna. Inst. fuer Sozio-Oekonomische Entwicklungsforschung
Description
The conventional method to evaluate the deadtime and countrate losses of scintillation counters often overestimates these performance parameters. This is caused by the fact that no countrate can be measured without losses. Therefore expected countrates are incorrect. This overestimation of system performance is of some importance for the clinical use of cameras, as scattered radiation from the patient decreases the ideal values deduced from measurements in air. Correct performance parameters can be calculated from 'relative countrates', i.e. the ratio of measured countrate over activity in arbitrary units. This ratio is drawn graphically on a logarithmic scale against activity. The monotonous curve yields the deadtime and 10% (20%) countrate loss unambiguously. The evaluation is independant of any deadtime concept or of the individual countrate loss or defective measurements. (orig.)
Abstract (German)
Bei der konventionellen Auswertung der Messdaten werden Totzeit und 10%- bzw. 20%-Zaehlratenverlust zu optimistisch ermittelt. Grund dafuer ist die Tatsache, dass auch der niedrigste Messwert nicht ohne Zaehlverlust gemessen werden kann. Durch die Berechnung relativer Zaehlraten - Quotient aus gemessener Zaehlrate durch Aktivitaet in willkuerlichen Einheiten - werden diese Parameter fehlerfrei bestimmt. Die relativen Zaehlraten werden halblogarithmisch aufgezeichnet. Ausreisser koennen dadurch eliminiert werden. Die Auswertung ist unabhaengig davon, wie hoch die Zaehlverluste der Messwerte sind und ob Modellvorstellungen, z.B. fuer die paralysierbare Totzeit, gelten oder nicht. (orig.)Additional details
Additional titles
- Original title (German)
- Eine einfache graphische Methode zur Ermittlung von Totzeit und Zaehlverlusten
Publishing Information
- Journal Title
- Nuc Compact
- Journal Volume
- 13
- Journal Issue
- 2
- Series
- Nuc Compact.
- Journal Page Range
- 68-72
- ISSN
- 0344-3752
INIS
- Country of Publication
- Germany
- Country of Input or Organization
- Germany
- INIS RN
- 13699803
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ALGORITHMS; COUNTING RATES; DEAD TIME; DECAY; GAMMA CAMERAS; HALF-LIFE; ISOMERIC NUCLEI; QUALITY CONTROL; RADIOACTIVITY; TECHNETIUM 99
- Descriptors DEC
- BETA DECAY RADIOISOTOPES; BETA-MINUS DECAY RADIOISOTOPES; CAMERAS; CONTROL; HOURS LIVING RADIOISOTOPES; INTERMEDIATE MASS NUCLEI; ISOMERIC TRANSITION ISOTOPES; ISOTOPES; NUCLEI; ODD-EVEN NUCLEI; RADIOISOTOPES; TECHNETIUM ISOTOPES; TIMING PROPERTIES; YEARS LIVING RADIOISOTOPES