Published September 1985
| Version v1
Journal article
Zenit-100 low-background noise installation for measuring beta radiation spatial distribution
- 1. Joint Inst. for Nuclear Research, Dubna (USSR)
Description
Low-background scanning installation for investigation of beta radiation distribution at the mass-separator collector is described. A detecting part of the device consists of 100 SI - 19 BG geiger counters, connected in the linear block with 100 cm length. Active and passive shielding decreases the background from external radiations about 6.5 times. Background for one counter is 0.002 s-1. The specific design collector provides spatial resolution of 3.3 mm
Additional details
Additional titles
- Original title (Russian)
- Низкофоновая установка Зенит-100 для измерения пространственног го распределения бета-активности
Publishing Information
- Journal Title
- Prib. Tekh. Ehksp.
- Journal Issue
- no.5
- Series
- Prib. Tekh. Ehksp.
- ISSN
- 0032-8162
- CODEN
- PRTEA
INIS
- Country of Publication
- Russian Federation
- Country of Input or Organization
- USSR
- INIS RN
- 17089773
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- BACKGROUND NOISE; BETA DETECTION; BETA SPECTROMETERS; DIAGRAMS; EFFICIENCY; GEIGER-MUELLER COUNTERS; MASS SPECTRA; MULTI-CHANNEL ANALYZERS; SHIELDING; SPATIAL DISTRIBUTION
- Descriptors DEC
- CHARGED PARTICLE DETECTION; DETECTION; DISTRIBUTION; ELECTRONIC EQUIPMENT; EQUIPMENT; INFORMATION; MEASURING INSTRUMENTS; NOISE; PULSE ANALYZERS; RADIATION DETECTION; RADIATION DETECTORS; SPECTRA; SPECTROMETERS
Optional Information
- Notes
- For English translation see the journal Instruments and Experimental Techniques (USA).