Published March 2012 | Version v1
Journal article

Evaluation of a new computerized analysis system developed for the processing of CT follow-up scans after EVR of infrarenal aneurysm

  • 1. Einzelunternehmen Dr. Julian Mattes and Software Competence Center Hagenberg GmbH, Softwarepark 35, Hagenberg (Austria)
  • 2. Department of Radiology, Innsbruck Medical University, Innsbruck (Austria)
  • 3. UMIT, Private University for Public Health, Medical Decision Making and Health Technology Assessment, Innsbruck (Austria)

Description

Purpose: The aim of this retrospective study was to present a new computerized analysis system developed for the evaluation of follow-up CT scans after endovascular repair (EVR) of infrarenal aneurysm and to compare it to the conventional evaluation method as regards preciseness and ease of application. The system is based on the extraction of the surface of the stent-graft (SG) and that of the spinal canal and the overlay of surfaces obtained at different points in time. Materials and methods: A total of 116 CT follow-up data sets obtained from 49 patients after EVR of infrarenal aneurysm were evaluated using both, the conventional method and the new computerized system. Two parameters were analyzed: SG length and the distance between the most ventral point of the SG and the vertebral column. The correlation between the results of the two methods and the correlation between the results obtained by two independent observers (radiologist and lay person) using the new system were assessed by statistical analysis. Results: Comparison of the two methods yielded a very high correlation for both parameters, (correlation coefficients of around 0.9 and p < 0.001). Comparison of the results obtained by the two observers yielded an equally high correlation (correlation coefficients of around 0.9 and p < 0.001). Conclusion: Our results show that the new computerized system is as precise and reliable as the conventional method, but allows better visualization and quantification of SG changes by surface overlay. Moreover, it is easier to apply and less time-consuming and can be easily integrated into existing systems.

Availability note (English)

Available from http://dx.doi.org/10.1016/j.ejrad.2010.12.070

Additional details

Identifiers

DOI
10.1016/j.ejrad.2010.12.070;
PII
S0720-048X(11)00019-2;

Publishing Information

Journal Title
European Journal of Radiology
Journal Volume
81
Journal Issue
3
Journal Page Range
p. 496-501
ISSN
0720-048X
CODEN
EJRADR

INIS

Country of Publication
Netherlands
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
43078475
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
CAT SCANNING; COMPARATIVE EVALUATIONS; GRAFTS; KIDNEYS; PATIENTS
Descriptors DEC
BODY; COMPUTERIZED TOMOGRAPHY; DIAGNOSTIC TECHNIQUES; EVALUATION; ORGANS; TOMOGRAPHY; TRANSPLANTS

Optional Information

Copyright
Copyright (c) 2011 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.