Published 1989 | Version v1
Report Open

Device for automatic measurement of dose fields

Description

A device is developed for measuring the dose characteristics of charged particle beams. Semiconductor diodes having sensitive volume from 0.1 to 10 mm3 are used as detectors. Electromechanical construction provides detector displacement in water or air medium in three perpendicular directions independently by computer or in hand regime. Coordinate measuring was made in a digital manner using the method of sequential calculation. Accuracy of calculating the location of transducer in any of three directions is 0.2 mm, limits and measuring time along X and Y axes were 0-250 mm in 60 and 80 s, along Z axis - 0-160 mm in 6 min. The detector and monitor reading values are introduced into computer on each cycle of accelerator work and normalized by program manager. The program was created for constructing the isodose maps over the results of measuring the relative dose values in the given points of medium. 4 refs.; 7 figs

Availability note (English)

MF available from INIS under the Report Number.

Files

22056227.pdf

Files (187.8 kB)

Name Size Download all
md5:9a44f021c3f71a23d2ed0205b4960e33
187.8 kB Preview Download

System files (19.3 kB)

Name Size Download all

Additional details

Additional titles

Original title (Russian)
Устройство для автоматизированного измерения дозных полей

Publishing Information

Imprint Pagination
5 p.
Report number
JINR-R--10-89-194