Published January 22, 1994 | Version v1
Journal article

Radioactivity measurements of 192Ir wire sources with a microcalorimeter

  • 1. Japan Atomic Energy Research Inst., Ibaraki (Japan)
  • 2. Dept. of Nuclear Engineering, Nagoya Univ. (Japan)

Description

Radioactivities of metallic 192Ir sources are measured with a microcalorimeter nondestructively. In order to convert the complete decay energy evolved from the source into thermal energy, a tungsten block is used as radiation absorber. The sample sources measured are four thin wires of Ir/Pt alloy with dimensions of 0.1 mmφ x 10 mm1, each coated by 0.1 mm thick platinum, giving the final diameter of 0.3 mmφ. Overall uncertainty of the measurement for some 60 MBq activity is ±3.2%. Measurements of the wire sources with an ionisation chamber give smaller activities which are smaller by 10% than those from microcalorimetry mainly because of the autoattenuation effect. (orig.)

Additional details

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
339
Journal Issue
1-2
Journal Page Range
p. 398-401.
ISSN
0168-9002
CODEN
NIMAER

Conference

Title
Conference of the International Committee for Radionuclide Metrology (ICRM) and international symposium on radionuclide metrology and its applications.
Dates
07-11 Jun 1993.
Place
Teddington (United Kingdom).