Published December 1, 1974 | Version v1
Journal article

Intensity measurement with a fixed specimen in a neutron single-crystal diffractometer

Creators

  • 1. Wuerzburg Univ. (F.R. Germany). Physikalisches Inst.

Description

In neutron single-crystal diffractometry integrated intensities are usually measured by rotating the specimen through the reflecting position. It is possible, however, to produce a monochromatic beam with properties that allow intensity measurement with a fixed specimen. For this bent monochromators are especially suitable, but plane mosaic-crystal monochromators in antiparallel (+) position can also be used. The production, suitability and use of a plastically bent germanium monochromator is discussed. By means of a structure determination of the room-temperature phase of NH4Br, intensity measurements with a fixed specimen are compared with measurements of integrated intensities. With regard to accuracy, the methods are equivalent. With a bent germanium monochromator, the intensity measurement with a fixed specimen leads, especially in the range of high Bragg angles, to a reduction of the measuring time by a factor of at least two.

Additional details

Additional titles

Original title (German)
Intensitaetsmessung bei ruhender Probe am Neutronen-Einkristalldiffraktometer

Identifiers

Publishing Information

Journal Title
Journal of Applied Crystallography
Journal Volume
7
Journal Issue
6
Series
J. Appl. Crystallogr.
Journal Page Range
564-572
ISSN
0021-8898

Optional Information

Notes
Updated automatically by Metadata and Full-Text Enrichment Agent