Error analysis of the microradiographical determination of mineral content in mineralised tissue slices
Creators
- 1. Rijksuniversiteit Groningen (Netherlands). Dental School
Description
The microradiographic method, used to measure the mineral content in slices of mineralised tissues as a function of position, is analysed. The total error in the measured mineral content is split into systematic errors per microradiogram and random noise errors. These errors are measured quantitatively. Predominant contributions to systematic errors appear to be x-ray beam inhomogeneity, the determination of the step wedge thickness and stray light in the densitometer microscope, while noise errors are under the influence of the choice of film, the value of the optical film transmission of the microradiographic image and the area of the densitometer window. Optimisation criteria are given. The authors used these criteria, together with the requirement that the method be fast and easy to build an optimised microradiographic system. (author)
Additional details
Publishing Information
- Journal Title
- Phys. Med. Biol.
- Journal Volume
- 30
- Journal Issue
- 10
- Series
- Phys. Med. Biol.
- Journal Page Range
- 1067-1075
- ISSN
- 0031-9155
- CODEN
- PHMBA
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- United Kingdom
- INIS RN
- 17009573
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- BONE TISSUES; ERRORS; MICRORADIOGRAPHY; MINERALS; NOISE; PHOTON BEAMS
- Descriptors DEC
- BEAMS; BODY; CONNECTIVE TISSUE; TISSUES