Published October 1985 | Version v1
Journal article

Error analysis of the microradiographical determination of mineral content in mineralised tissue slices

  • 1. Rijksuniversiteit Groningen (Netherlands). Dental School

Description

The microradiographic method, used to measure the mineral content in slices of mineralised tissues as a function of position, is analysed. The total error in the measured mineral content is split into systematic errors per microradiogram and random noise errors. These errors are measured quantitatively. Predominant contributions to systematic errors appear to be x-ray beam inhomogeneity, the determination of the step wedge thickness and stray light in the densitometer microscope, while noise errors are under the influence of the choice of film, the value of the optical film transmission of the microradiographic image and the area of the densitometer window. Optimisation criteria are given. The authors used these criteria, together with the requirement that the method be fast and easy to build an optimised microradiographic system. (author)

Additional details

Publishing Information

Journal Title
Phys. Med. Biol.
Journal Volume
30
Journal Issue
10
Series
Phys. Med. Biol.
Journal Page Range
1067-1075
ISSN
0031-9155
CODEN
PHMBA

INIS

Country of Publication
United Kingdom
Country of Input or Organization
United Kingdom
INIS RN
17009573
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
BONE TISSUES; ERRORS; MICRORADIOGRAPHY; MINERALS; NOISE; PHOTON BEAMS
Descriptors DEC
BEAMS; BODY; CONNECTIVE TISSUE; TISSUES