Published December 1984 | Version v1
Journal article

Standardization of technetium-99 by liquid-scintillation counting

  • 1. National Bureau of Standards, Washington, DC (USA). Radioactivity Group
  • 2. UKAEA Atomic Energy Research Establishment, Harwell. Environmental and Medical Sciences Div.
  • 3. Food and Drug Administration, Washington, DC (USA)
  • 4. Food and Drug Administration, Rockville, MD (USA)

Description

Technetium-99 has been standardized by comparing its pulse-height response on a liquid-scintillation counter with that of another β-particle-emitting standard radionuclide. In this work, hydrogen-3, carbon-14, and cobalt-60 were used as the standards, and the results obtained agreed to within 0.32%. The mass of potassium pertechnetate was also measured by gravimetric techniques for the technetium-99 radioactivity standard solution. A half-life value of (2.111 +- 0.012) x 105 years is suggested. The estimated uncertainty is intended to approximate one standard deviation. (author)

Additional details

Publishing Information

Journal Title
Int. J. Appl. Radiat. Isot.
Journal Volume
35
Journal Issue
12
Series
Int. J. Appl. Radiat. Isot.
Journal Page Range
1103-1112
ISSN
0020-708X