Ionization chamber dosimetry at a linear acceleration used for radiation therapy: analysis of dose deviations stemming from the use of different dosimetry concepts
Description
At a linear accelerator used for medical purposes, a cylindrical and a flat ionization chamber, respectively, were calibrated using the procedures: (1) Water energy (deposition) calibration, (2) Standard ion dose (=air) calibration, (3) Markus chamber calibration, all with regard to water energy dose which is the relevant quantity for radiation therapy planning. The primary accelerator energy range was 4-25 meV, and depth dependent dose rates for photons and electrons were considered. The aim is the comparison of water energy doses obtained by the three calibration methods, and an estimate of the resulting uncertainty of each, which turns out to be about 5%. (Quittner)
Availability note (English)
Available from the Technische Universitaet Graz, Petersgasse 16, A-8010 Graz (AT).Additional details
Additional titles
- Original title (German)
- Ionisationskammerdosimetrie am medizinisch genutzten Linearbeschleuniger
- Original subtitle (German)
- Analyse von Dosisabweichungen durch die Verwendung verschiedener Dosimetriekonzepte.
Publishing Information
- Imprint Pagination
- 85 p.
INIS
- Country of Publication
- Austria
- Country of Input or Organization
- Austria
- INIS RN
- 25022870
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Thesis, Non-conventional Literature
- Descriptors DEI
- ACCURACY; CALIBRATION; COMPARATIVE EVALUATIONS; DEPTH DOSE DISTRIBUTIONS; ELECTRON DOSIMETRY; GAMMA DOSIMETRY; IONIZATION CHAMBERS; MEV RANGE 01-10; MEV RANGE 10-100; RADIATION DOSES; RADIOTHERAPY
- Descriptors DEC
- DISTRIBUTION; DOSIMETRY; ENERGY RANGE; EVALUATION; MEASURING INSTRUMENTS; MEDICINE; MEV RANGE; RADIATION DETECTORS; RADIATION DOSE DISTRIBUTIONS; SPATIAL DISTRIBUTION; SPATIAL DOSE DISTRIBUTIONS; THERAPY
Optional Information
- Notes
- Referenz number II 74.776/A.