Published June 1990 | Version v1
Journal article

Basic imaging properties in a computed radiography system

  • 1. Yamaguchi Univ., Ube (Japan). School of Medicine

Description

Basic imaging properties in a computed radiography system (FRC-101), such as the dose absorption rate, relative intensities of instantaneous and photostimulated light, spatial resolution and noise properties, were measured. Three different versions of the photostimulable phosphor plate (image plate, IP), types I, II and III of ST and types II and III of HR were investigated. The presampling MTFs as an inherent resolution property in a digital system were measured for resolution analysis. Digital Wiener spectra were calculated for noise analysis. Moreover, the observer performance experiments with a Burger phantom, for types II and III of ST, were examined. Results showed that with the improvement of IPs, the instantaneous emission decreases, while the photostimulated emission increases considerably (approximately 1.8 and 1.7 times for ST and HR, respectively). A comparison of the different versions of IP showed an insignificant change in spatial-resolution properties but a significant improvement in noise. The detectability of low-contrast objects in type III was slightly better than that in type II. (author)

Additional details

Publishing Information

Journal Title
Nippon Hoshasen Gijutsu Gakkai Zasshi
Journal Volume
46
Journal Issue
6
Series
Nippon Hoshasen Gijutsu Gakkai Zasshi.
Journal Page Range
824-830
ISSN
0369-4305
CODEN
NIPHA