Phototimer for automatic control of x-ray exposure, PT-20A
- 1. Toshiba Corp., Kawasaki, Kanagawa (Japan)
Description
The phototimer as an exposure control device is now required to have higher accuracy and wider range of control to comply with the recent trends of X-ray diagnosis, such as toward short-time exposure due to increased output power of X-ray apparatus and toward one-man operation. In order to meet the requirement, the X-ray characteristics of the components of the phototimer have been examined, and computer simulation of X-ray spectra and dynamic characteristics of every component have been investigated, in pursuit of the essential characters of phototimers. On the basis of these analytical investigations, the phototimer PT-20A is so designed as to compensate for density deviation under various kVp and object thickness: uniform density has been obtained under practical radiographic conditions. (author)