Dosimetric characteristics of a multileaf collimator

  • 1. Turin Ospedale Mauriziano Umberto I, Turin (Italy). Unita' Operativa di fisica sanitaria

Description

It has been studied the dosimetric characteristics of a multileaf collimator (MLC) installed on a dual energy accelerator with 6 and 18 MV photon beams in the radiotherapy department of Mauriziano Umberto I Hospital in Turin (Italy) initiating its use in clinical practice. In particular, measurements included transmission through and between the leaves and at the junction under closed-leaves, central axis percentage depth dose, output factors and effective penumbra

Abstract (Italian)

Vengono analizzate e valutate le caratteristiche dosimetriche di un collimatore multilamellare installato su un acceleratore lineare Varian Clinac 2100 C/D con fotoni di due energie (6 e 18 MV) presente nell'Unita' Operativa di Radioterapia dell'Ospedale Mauriziano Umberto I di Torino. In particolare e' stata misurata la trasmissione sotto e tra lamelle, la fuga della radiazione alla giunzione tra lamelle opposte, la percentuale relativa della dose in profondita', i fattori di campo e la penombra effettiva

Additional details

Additional titles

Original title (Italian)
Caratterizzazione dosimetrica di un collimatore multilamellare

Publishing Information

Journal Title
Radiologia Medica
Journal Volume
97
Journal Issue
5
Journal Page Range
p. 382-388
ISSN
0033-8362
CODEN
RAMEAN

INIS

Country of Publication
Italy
Country of Input or Organization
Italy
INIS RN
31021560
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
COLLIMATORS; DOSIMETRY; NUCLEAR MEDICINE; RADIOTHERAPY; TOMOGRAPHY
Descriptors DEC
MEDICINE; THERAPY