Published May 1999
| Version v1
Journal article
Dosimetric characteristics of a multileaf collimator
- 1. Turin Ospedale Mauriziano Umberto I, Turin (Italy). Unita' Operativa di fisica sanitaria
Description
It has been studied the dosimetric characteristics of a multileaf collimator (MLC) installed on a dual energy accelerator with 6 and 18 MV photon beams in the radiotherapy department of Mauriziano Umberto I Hospital in Turin (Italy) initiating its use in clinical practice. In particular, measurements included transmission through and between the leaves and at the junction under closed-leaves, central axis percentage depth dose, output factors and effective penumbra
Abstract (Italian)
Vengono analizzate e valutate le caratteristiche dosimetriche di un collimatore multilamellare installato su un acceleratore lineare Varian Clinac 2100 C/D con fotoni di due energie (6 e 18 MV) presente nell'Unita' Operativa di Radioterapia dell'Ospedale Mauriziano Umberto I di Torino. In particolare e' stata misurata la trasmissione sotto e tra lamelle, la fuga della radiazione alla giunzione tra lamelle opposte, la percentuale relativa della dose in profondita', i fattori di campo e la penombra effettivaAdditional details
Additional titles
- Original title (Italian)
- Caratterizzazione dosimetrica di un collimatore multilamellare
Publishing Information
- Journal Title
- Radiologia Medica
- Journal Volume
- 97
- Journal Issue
- 5
- Journal Page Range
- p. 382-388
- ISSN
- 0033-8362
- CODEN
- RAMEAN
INIS
- Country of Publication
- Italy
- Country of Input or Organization
- Italy
- INIS RN
- 31021560
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- COLLIMATORS; DOSIMETRY; NUCLEAR MEDICINE; RADIOTHERAPY; TOMOGRAPHY
- Descriptors DEC
- MEDICINE; THERAPY