Method validation to determine total alpha beta emitters in water samples using LSC
- 1. Atomic Energy Commission, Damascus (Syrian Arab Republic), Dept. of Protection and Safety
Description
In this work a method was validated to determine gross alpha and beta emitters in water samples using liquid scintillation counter. 200 ml of water from each sample were evaporated to 20 ml and 8 ml of them were mixed with 12 ml of the suitable cocktail to be measured by liquid scintillation counter Wallac Winspectral 1414. The lower detection limit by this method (LDL) was 0.33 DPM for total alpha emitters and 1.3 DPM for total beta emitters. and the reproducibility limit was (+- 2.32 DPM) and (+-1.41 DPM) for total alpha and beta emitters respectively, and the repeatability limit was (+-2.19 DPM) and (+-1.11 DPM) for total alpha and beta emitters respectively. The method is easy and fast because of the simple preparation steps and the large number of samples that can be measured at the same time. In addition, many real samples and standard samples were analyzed by the method and showed accurate results so it was concluded that the method can be used with various water samples. (author)
Additional details
Publishing Information
- Journal Title
- Aalam Al-Zarra
- Journal Issue
- 112
- Journal Page Range
- p. 63
- ISSN
- 1607-985X
- CODEN
- AAALE5
INIS
- Country of Publication
- Syrian Arab Republic
- Country of Input or Organization
- Syrian Arab Republic
- INIS RN
- 38114394
- Subject category
- S54: ENVIRONMENTAL SCIENCES;
- Descriptors DEI
- ACCURACY; ALPHA DETECTION; BETA DETECTION; CALIBRATION; LIQUID SCINTILLATION DETECTORS; SAMPLE PREPARATION; SENSITIVITY; VALIDATION; WATER
- Descriptors DEC
- CHARGED PARTICLE DETECTION; DETECTION; HYDROGEN COMPOUNDS; MEASURING INSTRUMENTS; OXYGEN COMPOUNDS; RADIATION DETECTION; RADIATION DETECTORS; SCINTILLATION COUNTERS; TESTING
Optional Information
- Notes
- Abstract of Scientific Research