Real-time intrafraction motion monitoring in external beam radiotherapy
Creators
- 1. Joint Department of Physics, Institute of Cancer Research and Royal Marsden NHS Foundation Trust, London (United Kingdom)
- 2. Department of Radiation Oncology, University Medical Center Groningen, University of Groningen (Netherlands)
- 3. Department of Medical Physics and Biomedical Engineering, Centre for Medical Image Computing, University College London, London (United Kingdom)
- 4. Department of Oncology, Aarhus University Hospital, Aarhus (Denmark)
- 5. ACRF Image X Institute, University of Sydney, Sydney (Australia)
Description
Radiotherapy (RT) aims to deliver a spatially conformal dose of radiation to tumours while maximizing the dose sparing to healthy tissues. However, the internal patient anatomy is constantly moving due to respiratory, cardiac, gastrointestinal and urinary activity. The long term goal of the RT community to 'see what we treat, as we treat' and to act on this information instantaneously has resulted in rapid technological innovation. Specialized treatment machines, such as robotic or gimbal-steered linear accelerators (linac) with in-room imaging suites, have been developed specifically for real-time treatment adaptation. Additional equipment, such as stereoscopic kilovoltage (kV) imaging, ultrasound transducers and electromagnetic transponders, has been developed for intrafraction motion monitoring on conventional linacs. Magnetic resonance imaging (MRI) has been integrated with cobalt treatment units and more recently with linacs. In addition to hardware innovation, software development has played a substantial role in the development of motion monitoring methods based on respiratory motion surrogates and planar kV or Megavoltage (MV) imaging that is available on standard equipped linacs.
In this paper, we review and compare the different intrafraction motion monitoring methods proposed in the literature and demonstrated in real-time on clinical data as well as their possible future developments. We then discuss general considerations on validation and quality assurance for clinical implementation.
Besides photon RT, particle therapy is increasingly used to treat moving targets. However, transferring motion monitoring technologies from linacs to particle beam lines presents substantial challenges. Lessons learned from the implementation of real-time intrafraction monitoring for photon RT will be used as a basis to discuss the implementation of these methods for particle RT. (topical review)
Availability note (English)
Available from http://dx.doi.org/10.1088/1361-6560/ab2ba8Additional details
Identifiers
Publishing Information
- Journal Title
- Physics in Medicine and Biology
- Journal Volume
- 64
- Journal Issue
- 15
- Journal Page Range
- [33 p.]
- ISSN
- 0031-9155
- CODEN
- PHMBA7
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 52003963
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- COMPUTER CODES; EQUIPMENT; EXTERNAL BEAM RADIATION THERAPY; LINEAR ACCELERATORS; NMR IMAGING; PARTICLE BEAMS; QUALITY ASSURANCE; RADIATION DOSES
- Descriptors DEC
- ACCELERATORS; BEAMS; DIAGNOSTIC TECHNIQUES; DOSES; MANAGEMENT; MEDICINE; NUCLEAR MEDICINE; QUALITY MANAGEMENT; RADIOLOGY; RADIOTHERAPY; THERAPY