Published 1977
| Version v1
Report
Pair γ spectrometer
Description
A paring gamma spectrometer is described with the resolution of 35 keV at Esub(γ)=10 MeV developed on the base of a germanium detector and a four-section NaI (Tl) scintillator. Efficiency of the spectrometer is 10% of efficiency of detecting with the single-crystal method and is equal to 0.03% in the energy range Esub(γ) from 4 to 12 MeV. Spectrometer is aimed at studying (p, γ) reactions at low and medium proton energies
Additional details
Additional titles
- Original title (Russian)
- Парный γ-спектрометр
Publishing Information
- Imprint Title
- High energy physics and nuclear physics
- Journal Issue
- no. 1
- Series
- Voprosy atomnoj nauki i tekhniki.
- Journal Page Range
- p. 50-51.
- Report number
- KFTI--77-7
Conference
- Title
- Kharkov Physico-Technical Institute summary conference on nuclear physics and elementary particle physics.
- Dates
- 1977.
- Place
- Kharkov, Ukrainian SSR.
INIS
- Country of Publication
- USSR
- Country of Input or Organization
- USSR
- INIS RN
- 10438210
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- ACCURACY; COINCIDENCE CIRCUITS; DISCRIMINATORS; ENERGY RESOLUTION; GAMMA SPECTRA; INTEGRATED CIRCUITS; LI-DRIFTED GE DETECTORS; MEV RANGE 01-10; OPERATIONAL AMPLIFIERS; PAIR SPECTROMETERS; PROTON BEAMS; SOLID SCINTILLATION DETECTORS; TIME RESOLUTION
- Descriptors DEC
- AMPLIFIERS; BEAMS; ELECTRONIC CIRCUITS; ENERGY RANGE; GAMMA SPECTROMETERS; GE SEMICONDUCTOR DETECTORS; LI-DRIFTED DETECTORS; MEASURING INSTRUMENTS; MEV RANGE; NUCLEON BEAMS; PARTICLE BEAMS; RADIATION DETECTORS; RESOLUTION; SCINTILLATION COUNTERS; SEMICONDUCTOR DETECTORS; SPECTRA; SPECTROMETERS; TIMING PROPERTIES
Optional Information
- Notes
- 3 refs.; 3 figs. Imprint:Fizika vysokikh ehnergij i atomnogo yadra.