Published November 2008 | Version v1
Journal article

Modeling of carbon fiber couch attenuation properties with a commercial treatment planning system

  • 1. Department of Radiation Oncology, University of Arkansas for Medical Sciences, 4301 W. Markham Street, Little Rock, Arkansas 72205 (United States)

Description

The purpose of this work is to evaluate the modeling of carbon fiber couch attenuation properties with a commercial treatment planning system (TPS, Pinnacle3, v8.0d). A carbon fiber couch (BrainLab) was incorporated into the TPS by automatic contouring of all transverse CT slices. The couch shape and dimensions were set according to the vendor specifications. The couch composition was realized by assigning appropriate densities to the delineated contours. The couch modeling by the TPS was validated by absolute dosimetric measurements. A phantom consisting of several solid water slabs was CT scanned, the CT data set was imported into the TPS, and the carbon fiber couch was auto-contoured. Open (unblocked) field plans for different gantry angles and field sizes were generated. The doses to a point at 3 cm depth, placed at the linac isocenter, were computed. The phantom was irradiated according to the dose calculation setup and doses were measured with an ion chamber. In addition, percent depth dose (PDD) curves were computed as well as measured with radiographic film. The calculated and measured doses, transmissions, and PDDs were cross-compared. Doses for several posterior fields (0 deg., 30 deg., 50 deg., 75 deg., 83 deg.) were calculated for 6 and 18 MV photon beams. For model validation a nominal field size of 10x10 cm2 was chosen and 100 MU were delivered for each portal. The largest difference between computed and measured doses for those posterior fields was within 1.7%. A comparison between computed and measured transmissions for the aforementioned fields was performed and the results were found to agree within 1.1%. The differences between computed and measured doses for different field sizes, ranging from 5x5 cm2 to 25x25 cm2 in 5 cm increments, were within 2%. Measured and computed PDD curves with and without the couch agree from the surface up to 30 cm depth. The PDDs indicate a surface dose increase resulting from the carbon fiber couch field modification. The carbon fiber couch attenuation for individual posterior oblique fields (75 deg.) can be in excess of 8% depending on the beam energy and field size. When the couch is contoured in Pinnacle3 its attenuation properties are modeled to within 1.7% with respect to measurements. These results demonstrate that appropriate contouring together with relevant density information for the contours is sufficient for adequate modeling of carbon fiber supporting devices by modern commercial treatment planning systems.

Additional details

Identifiers

Publishing Information

Journal Title
Medical Physics
Journal Volume
35
Journal Issue
11
Journal Page Range
p. 4982-4988
ISSN
0094-2405
CODEN
MPHYA6

Optional Information

Notes
(c) 2008 American Association of Physicists in Medicine