Published February 1, 1996 | Version v1
Journal article

Reduction of γ-ray background in the use of 32 keV X-rays from 137Cs sources by a gating method with a thin plastic scintillation detector

  • 1. Seikei Univ., Tokyo (Japan). Fac. of Eng.
  • 2. Electrotechnical Lab., Thukuba, Ibaraki (Japan)

Description

In the use of 32 keV K X-rays from 137Cs sources for the calibration of low energy photon spectrometers, a new method was developed to reduce remarkably the background continuum due to 662 keV γ-rays. This method is based upon a gating technique by the use of a 4π detector composed of thin plastic scintillator sheets which sandwich the 137Cs source. Since all X-rays from the 137Cs source are coincident with monoenergetic internal conversion electrons, the X-ray response of the photon spectrometer remains unchanged, whereas γ-ray signals are rejected by gating. This method was tested successfully for three different photon spectrometers. (orig.)

Additional details

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
369
Journal Issue
2-3
Journal Page Range
p. 671-675.
ISSN
0168-9002
CODEN
NIMAER

Conference

Title
International symposium on radionuclide metrology and its applications.
Original Conference Title
Symposium International sur la Metrologie des Radionucleides et ses Applications
Dates
15-19 May 1995.
Place
Paris (France).