Published April 7, 2005 | Version v1
Journal article

The use of detective quantum efficiency (DQE) in evaluating the performance of gamma camera systems

  • 1. Department of Radiation Physics, Goeteborg University, Goeteborg (Sweden)
  • 2. Department of Hospital Physics, County Hospital Ryhov, Joenkoeping (Sweden)
  • 3. Department of Medical Physics and Biomedical Engineering, Sahlgrenska University Hospital, Goeteborg (Sweden)
  • 4. Department of Radiology, Uddevalla Hospital, Uddevalla (Sweden)

Description

The imaging properties of an imaging system can be described by its detective quantum efficiency (DQE). Using the modulation transfer function calculated from measured line spread functions and the normalized noise power spectrum calculated from uniformity images, DQE was calculated with the number of photons emitted from a plane source as a measure for the incoming SNR2. Measurements were made with 99mTc, using three different pulse height windows at 2 cm and 12 cm depths in water with high resolution and all purpose collimators and with two different crystal thicknesses. The results indicated that at greater depths a 15% window is the best choice. The choice of collimator depends on the details in the organ being investigated. There is a break point at 0.5 cycles cm-1 and 1.2 cycles cm-1 at 12 cm and 2 cm depths, respectively. A difference was found in DQE between the two crystal thicknesses, with a slightly better result for the thick crystal for measurements at 12 cm depth. At 2 cm depth, the thinner crystal was slightly better for frequencies over 0.5 cm-1. The determination of DQE could be a method to optimize the parameters for different nuclear medicine investigations. The DQE could also be used in comparing different gamma camera systems with different collimators to obtain a figure of merit

Availability note (English)

Available online at http://stacks.iop.org/0031-9155/50/1601/pmb5_7_019.pdf or at the Web site for the journal Physics in Medicine and Biology (ISSN 1361-6560) http://www.iop.org/

Additional details

Publishing Information

Journal Title
Physics in Medicine and Biology
Journal Volume
50
Journal Issue
7
Journal Page Range
p. 1601-1609
ISSN
0031-9155
CODEN
PHMBA7