System for measurement of low yield neutron pulses from D–D fusion reactions based upon a 3He proportional counter

  • 1. Comisión Chilena de Energía Nuclear, Casilla 188-D, Santiago (Chile)
  • 2. Comisión Nacional de Energía Atómica, Univ. N. de Cuyo (Argentina)

Description

A conventional neutron detection technique was adapted to measure low neutron yields from D–D fusion pulses. This method uses a 3He proportional counter surrounded by a paraffin moderator. Electric signals generated in the 3He tube are fed into a preamplifier. The output of the preamplifier is directly connected to a digital oscilloscope. The time-integrated signals represent the charge generated in the 3He tube which is proportional to the total neutron yield. The integration time is determined by the preamplifier and moderator characteristics within some hundreds of microseconds. No meaningful neutron background was detected during this time window. The system, previously calibrated, was used to measure the neutron yield (<106 neutron/pulse) generated in a fast and very small plasma focus device designed to operate with energies of tens of Joules. Neutron yields as low as 103 neutrons per pulse were measured. (technical design note)

Availability note (English)

Available from http://dx.doi.org/10.1088/0957-0233/19/8/087002

Additional details

Identifiers

DOI
10.1088/0957-0233/19/8/087002;
PII
S0957-0233(08)62013-5;

Publishing Information

Journal Title
Measurement Science and Technology
Journal Volume
19
Journal Issue
8
Journal Page Range
[5 p.]
ISSN
0957-0233
CODEN
MSTCEP