Phase calibration of hydrophones: heterodyne time-delay spectrometry and broadband pulse technique using an optical reference hydrophone
Creators
- 1. Physikalisch-Technische Bundesanstalt, Bundesallee 100, 38116 Braunschweig (Germany)
Description
Two hydrophone calibration techniques are presented which provide amplitude and phase of hydrophone sensitivity. Heterodyne time-delay spectrometry can be applied in the frequency range from 0.5 to 50 MHz to all common hydrophones but needs a standard to obtain absolute values. A pulse technique between 1 and 70 MHz allows exploiting of an optical multilayer hydrophone as a phase standard. In combination with HTDS, a calibration service is established that covers the complete spectrum of common piezoelectric and optical hydrophones for absolute amplitude and phase calibration. The measured complex frequency responses form the basis of deconvolution procedures suitable for the correction of measurements of, for example, the output of ultrasound diagnostic machines
Availability note (English)
Available online at http://stacks.iop.org/1742-6596/1/14/jpconf4_1_006.pdf or at the Web site for the Journal of Physics. Conference Series (Online) (ISSN 1742-6596) http://www.iop.org/Additional details
Identifiers
Publishing Information
- Journal Title
- Journal of Physics. Conference Series (Online)
- Journal Volume
- 1
- Journal Issue
- 1
- Journal Page Range
- p. 14-19
- ISSN
- 1742-6596
Conference
- Title
- International conference on advanced metrology for ultrasound in medicine 2004
- Acronym
- AMUM 2004
- Dates
- 27-28 Apr 2004
- Place
- Teddington, Middlesex (United Kingdom)
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 37003178
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- AMPLITUDES; CALIBRATION; MHZ RANGE 01-100; PIEZOELECTRICITY; PULSE TECHNIQUES; SENSITIVITY; SPECTROSCOPY; STANDARDS; TIME DELAY
- Descriptors DEC
- ELECTRICITY; FREQUENCY RANGE; MHZ RANGE