Published 1990 | Version v1
Report Open

Setting up and performance of a laser enhanced ionisation spectrometer

  • 1. Bhabha Atomic Research Centre, Bombay (India). Spectroscopy Div.

Description

A laser enhanced ionisation (LEI) spectrometer has been successfuly set up around an excimer laser pumped dye laser. The performance of the spectrometer has been tested by analysing sodium in water solutions. A straight line working curve has been obtained in the concentration range of 1-1000 ng/ml of Na. The effect of parameters like laser power, ion collector electrode voltage and the load resistance on LEI signals has been investigated. The spectrometer is useful not only for analytical purposes but also for laser spectroscopic studies of species formed in flames, study of phenomenon of combustion etc. (author). 1 tab ., 10 figs., 5 refs

Availability note (English)

MF available from INIS under the Report Number.

Files

22019710.pdf

Files (416.9 kB)

Name Size Download all
md5:161f6c21381d9e95624c9792e22ecfa5
416.9 kB Preview Download

Additional details

Additional titles

Augmented title (English)
Laser enhanced ionization spectrometer

Publishing Information

Imprint Pagination
22 p.
Report number
BARC--1510