Published 1976 | Version v1
Journal article

Adaptation of the industrial betatron produced in Poland for medical applications (preliminary dose measurements)

  • 1. Institute of Nuclear Research, Warsaw (Poland)

Description

Isodose distributions of X-ray beam, the percent depth dose along the central beam axis, transversial dose distributions and the heating characteristics of an industrial type - B30 S betatron designed and built by the Institute of Nuclear Research at Swierk is given. The measurements were carried out with the aid of an ionization chamber and a photographic technique in water and organic glass phantoms. The preliminary results call for some modification in the flattening filter design, since those used so far do not yield isodose distribution suitable for teletherapy. (author)

Additional details

Additional titles

Original title (Polish)
Adaptacja betatronu przemyslowego produkcji krajowej do celow medycznych - wstepne prace dozymetryczne

Publishing Information

Journal Title
Postepy Fiz. Med.
Journal Volume
11
Journal Issue
3
Series
Postepy Fiz. Med.
Journal Page Range
141-147