Published September 21, 2004 | Version v1
Journal article

Graded-gap AlxGa1-xAs analyzer of X-ray and alpha-particle spectra

Creators

Description

Graded-gap AlxGa1-xAs detectors and spectrum analyzers for X-ray energies in the range of 5-100 keV are proposed and tested. The charge collection efficiency of the AlxGa1-xAs detectors decreases by about 90% after alpha-particle irradiation corresponding to a fluence of 1011 cm-2

Additional details

Identifiers

DOI
10.1016/j.nima.2004.05.099;
PII
S0168900204011258;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
531
Journal Issue
1-2
Journal Page Range
p. 92-96
ISSN
0168-9002
CODEN
NIMAER

Conference

Title
5. international workshop on radiation imaging detectors
Dates
7-11 Sep 2003
Place
Riga (Latvia)

Optional Information

Copyright
Copyright (c) 2004 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.