Published December 1, 2012 | Version v1
Journal article

A detailed test of a BSO calorimeter with 100–800 MeV positrons

  • 1. Research Center for Electron Photon Science, Tohoku University, Sendai 982-0826 (Japan)
  • 2. Department of Physics, Yamagata University, Yamagata 990-8560 (Japan)

Description

A performance test has been made for an electro-magnetic calorimeter prototype comprised four identical BSO crystals arranged in a 2×2 matrix by utilizing a positron beam in the energy range from 100 to 800 MeV. The size of each crystal is 40×40×210 mm3. This is the world's largest BSO single crystal ever used as a photon detector. The obtained energy resolution is (σE/E)2=((1.71%±0.03%)/√(E))2+(1.12%±0.08%)2 at room temperature, where E is the incident positron energy given in GeV. A BGO calorimeter having the same geometry has been employed in the performance test for comparison. The basic scintillation characteristics of BSO crystals measured without the beam are also presented.

Availability note (English)

Available from http://dx.doi.org/10.1016/j.nima.2012.08.085

Additional details

Identifiers

DOI
10.1016/j.nima.2012.08.085;
PII
S0168-9002(12)00988-6;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
694
Journal Page Range
p. 348-360