Published February 8, 1993 | Version v1
Journal article

More about track etching statistics

Creators

  • 1. Strossmayer Univ., Osijek (Yugoslavia). Faculty of Education

Description

Statistical errors of the track etching method were investigated by using the LR-115 nuclear track detector and visual microscopic counting. In the experiment, four persons counted tracks of the same detector and the analysis of variance was applied to the respective means and standard deviations. Another experiment was performed with ten detectors exposed to air in the same room; after standard etching and counting, the relative error of the detector track density was obtained as 11.2%. Taking into account the errors of the detector sensitivity coefficient and background, the total relative error of the track etching method was 19.5% in case of measurements of the indoor radon concentration of 8.2 Bq m-3. (author) 4 refs.; 2 tabs

Additional details

Publishing Information

Journal Title
Journal of Radioanalytical and Nuclear Chemistry
Journal Volume
175
Journal Issue
2
Journal Page Range
p. 95-101.
ISSN
0236-5731
CODEN
JRNCDM

INIS