Detection and analysis of xenon isotopes for the comprehensive nuclear-test-ban treaty international monitoring system

Description

The use of the xenon isotopes for detection of nuclear explosions is of great interest for monitoring compliance with the comprehensive nuclear-test-ban treaty (CTBT). Recently, the automated radioxenon sampler-analyzer (ARSA) was tested at the Institute for Atmospheric Radioactivity (IAR) in Freiburg, Germany to ascertain its use for the CTBT by comparing its results to laboratory-based analyses, determining its detection sensitivity and analyzing its results in light of historical xenon isotope levels and known reactor operations in the area. Xe-133 was detected nearly every day throughout the test at activity concentrations ranging between approximately 0.1 mBq/m3 to as high as 120 mBq/m3. Xe-133m and 135Xe were also detected occasionally during the test at concentrations of less than 1 to a few mBq/m3

Additional details

Identifiers

PII
S0265931X0100042X;

Publishing Information

Journal Title
Journal of Environmental Radioactivity
Journal Volume
59
Journal Issue
2
Journal Page Range
p. 139-151
ISSN
0265-931X
CODEN
JERAEE

Optional Information

Copyright
Copyright (c) 2002 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.