Design and characterization of a movable emittance meter for low-energy electron beams
Creators
- Catani, Luciano1, 2, 3
- Chiadroni, Enrica1, 2, 3
- Cianchi, Alessandro1, 2, 3
- Tazzari, Sergio1, 2, 3
- Boscolo, Manuela1, 2, 3
- Castellano, Michele1, 2, 3
- Di Pirro, Giampiero1, 2, 3
- Ferrario, Massimo1, 2, 3
- Fusco, Valeria1, 2, 3
- Filippetto, Daniele1, 2, 3
- Palumbo, Luigi1, 2, 3
- Vaccarezza, Cristina1, 2, 3
- Vicario, Carlo1, 2, 3
- Ronsivalle, Concetta1, 2, 3
- 1. ENEA, CR Frascati, Via Enrico Fermi 45, I-00044 Frascati (Italy)
- 2. INFN-LNF, Via Enrico Fermi 40, I-00044 Frascati (Italy)
- 3. INFN-Roma Tor Vergata, Via della Ricerca Scientifica 1, I-00133 Rome (Italy)
Description
In order to characterize and optimize the beam emittance compensation scheme [L. Serafini and J. B. Rosenzweig, Phys. Rev. E 55, 7565 (1997)] of the Sorgente Pulsata Autoamplificata di Radiazione Coerente project [D. Alesini et al., Nucl. Instrum. Methods Phys. Res. A 507, 345 (2003)] high brightness preinjector a system to accurately measure the beam rms emittance evolution downstream of the rf gun has been developed. Since in a space charge dominated beam the quadrupole-scan method is not applicable, a movable emittance measurement device has been built based on the pepper-pot technique. The device consists of a double system of horizontal and vertical slit arrays and a downstream screen, all installed on a longitudinally movable support equipped with bellows and spanning the ≅1.2 m long drift space between gun and first accelerating section. The system allows the measuring of the beam rms emittance all along the spanned region so as to accurately reconstruct its evolution along the beam trajectory. More than a simple improvement over more conventional beam emittance measurement tools this device defines a new strategy for characterizing high performance photoinjectors as it allows a detailed analysis of the beam behavior over a section of the accelerator where crucial beam shaping takes place. Numerical simulations of the measurement, mainly based on PARMELA [J. Billen, PARMELA Report No. LA-UR 96-1835, 1996 (unpublished)], have been used to estimate the achievable accuracy and to optimize the experimental setup. Wake field effects induced by the beam propagation through the bellows have also been investigated with HOMDYN [M. Ferrario et al., LCLS Report No. SLAC-PUB 84000, 1999 (unpublished)]. A series of laboratory tests to evaluate its performance has been carried out at LNF in Frascati. The system was then moved to DESY Zeuthen and installed on the Photo Injector Test Facility PITZ, for further testing with beam. Design criteria and tests carried out to evaluate the device performance are discussed
Additional details
Identifiers
- DOI
- 10.1063/1.2336763;
Publishing Information
- Journal Title
- Review of Scientific Instruments
- Journal Volume
- 77
- Journal Issue
- 9
- Journal Page Range
- p. 093301-093301.7
- ISSN
- 0034-6748
- CODEN
- RSINAK
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 38026635
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S43: PARTICLE ACCELERATORS;
- Descriptors DEI
- BEAM EMITTANCE; BEAM SHAPING; BELLOWS; BRIGHTNESS; DESIGN; DESY; ELECTRON BEAMS; METERS; PERFORMANCE; PHOTOCATHODES; SPACE CHARGE; STANFORD LINEAR ACCELERATOR CENTER; TEST FACILITIES
- Descriptors DEC
- ACCELERATORS; BEAMS; CATHODES; CYCLIC ACCELERATORS; ELECTRODES; LEPTON BEAMS; MEASURING INSTRUMENTS; NATIONAL ORGANIZATIONS; OPTICAL PROPERTIES; PARTICLE BEAMS; PHYSICAL PROPERTIES; SYNCHROTRONS; US DOE; US ERDA; US ORGANIZATIONS
Optional Information
- Notes
- (c) 2006 American Institute of Physics