The role of phantom and treatment head generated bremsstrahlung in high-energy electron beam dosimetry
Creators
- 1. Department of Medical Radiation Physics, Karolinska Institutet and Stockholm University, Box 260, S-171 76 Stockholm (Sweden)
Description
An analytical expression has been derived for the phantom generated bremsstrahlung photons in plane-parallel monoenergetic electron beams normally incident on material of any atomic number (Be, H2O, Al, Cu and U). The expression is suitable for the energy range from 1 to 50 MeV and it is solely based on known scattering power and radiative and collision stopping power data for the material at the incident electron energy. The depth dose distribution due to the bremsstrahlung generated by the electrons in the phantom is derived by convolving the bremsstrahlung energy fluence produced in the phantom with a simple analytical energy deposition kernel. The kernel accounts for both electrons and photons set in motion by the bremsstrahlung photons. The energy loss by the primary electrons, the build-up of the electron fluence and the generation, attenuation and absorption of bremsstrahlung photons are all taken into account in the analytical formula. The longitudinal energy deposition kernel is derived analytically and it is consistent with both the classical biexponential relation describing the photon depth dose distribution and the exponential attenuation of the primary photons. For comparison Monte Carlo calculated energy deposition distributions using ITS3 code were used. Good agreement was found between the results with the analytical expression and the Monte Carlo calculation. For tissue equivalent materials, the maximum total energy deposition differs by less than 0.2% from Monte Carlo calculated dose distributions. The result can be used to estimate the depth dependence of phantom generated bremsstrahlung in different materials in therapeutic electron beams and the bremsstrahlung production in different electron absorbers such as scattering foils, transmission monitors and photon and electron collimators. By subtracting the phantom generated bremsstrahlung from the total bremsstrahlung background the photon contamination generated in the treatment head can be determined to allow accurate dosimetry of heavily photon contaminated electron beams. (author)
Additional details
Publishing Information
- Journal Title
- Physics in Medicine and Biology (Online)
- Journal Volume
- 41
- Journal Issue
- 12
- Journal Page Range
- p. 2657-2677
- ISSN
- 1361-6560
INIS
- Country of Publication
- International Atomic Energy Agency (IAEA)
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 31025334
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- BREMSSTRAHLUNG; ELECTRON DOSIMETRY; ENERGY DEPOSITION; MONTE CARLO METHOD; PHANTOMS; PHOTON BEAMS; RADIATION DOSE DISTRIBUTIONS; RADIOTHERAPY
- Descriptors DEC
- BEAMS; CALCULATION METHODS; DOSIMETRY; ELECTROMAGNETIC RADIATION; MEDICINE; MOCKUP; RADIATIONS; STRUCTURAL MODELS; THERAPY
Optional Information
- Notes
- Refs