Published February 1994 | Version v1
Journal article

Development of a uniform efficiency photon detector

  • 1. Indian Association for the the Cultivation of Science, Calcutta (India)
  • 2. Indian Association for the Cultivation of Science, Calcutta (India). Bose Lab.

Description

A uniform efficiency photon counter was developed by means of a plastic scintillator in conjunction with an aluminium filter. The efficiency was measured by using four radioactive sources, viz. 203Hg, 51Cr, 137Cs and 54Mn. The achievement of the uniform efficiency was ascertained by measuring the total (sum of coherent and incoherent) differential scattering cross section from Sn and Pb atoms, and comparing the results with calculated cross section values. (author)

Additional details

Publishing Information

Journal Title
Applied Radiation and Isotopes
Journal Volume
45
Journal Issue
2
Journal Page Range
p. 225-228.
ISSN
0969-8043
CODEN
ARISEF