Published 1981 | Version v1
Report Open

Determination of ultratrace concentrations of uranium and thorium in natural waters by x-ray fluorescence

Description

An x-ray fluorescence method for the simultaneous determination of uranium and thorium at the less than 1 ppM level in natural waters is described. Uranium and thorium are coprecipitated with an internal standard, yttrium, and incorporated into an iron-aluminum hydroxide carrier. The hydroxide precipitate is filtered, and the filter disk is analyzed by the energy-dispersive x-ray fluorescence technique. Matrix interferences caused by the presence of unpredictable anions and cations are compensated for by the internal standard. The U/Y and Th/Y ratios are linear over the 5- to 100-μg range of interest, and the detection limit of each element on the filter disk is 2 μg (3 sigma). Relative standard deviation was 17% at the 15-μg and 4% at the 100-μg level for thorium and 11% at the 11-μg and 2% at the 100-μg level for uranium. Analysis of spiked solutions showed a recovery of 19.6 +- 0.3 μg for uranium and 19.8 +- 0.3 μg for thorium at the 20-μg level, and the normal lower reporting limit is 5 μg. Fifty disks can be routinely measured during a normal working day

Availability note (English)

MF available from INIS under the Report Number; Available from NTIS., PC A02/MF A01.

Files

12626436.pdf

Files (147.7 kB)

Name Size Download all
md5:badc402927e39435da837796a04425f3
147.7 kB Preview Download

Additional details

Publishing Information

Imprint Pagination
9 p.
Report number
CONF-810647--6

Conference

Title
Conference on X- and gamma-ray sources and applications.
Dates
10 - 12 Jun 1981.
Place