Published June 1999 | Version v1
Journal article

Temperature dependence of the calibration factor of radon and radium determination in water samples by SSNTD

Description

The sensitivity of a 226Ra determination method of water samples by SSNTD was measured as a function of storage temperature during exposure. The method is based on an etched track type radon monitor, which is closed into a gas permeable foil and is immersed in the water sample. The sample is sealed in a glass vessel and stored for an exposure time of 10-30 days. The sensitivity increased more than a factor of two when the storage temperature was raised from 2 deg. C to 30 deg. C. Temperature dependence of the partition coefficient of radon between water and air provides explanation for this dependence. For practical radio- analytical application the temperature dependence of the calibration factor is given by fitting the sensitivity data obtained by measuring 226Ra standard solutions (in the activity concentration range of 0.1-48.5 kBq m-3) at different storage temperatures

Additional details

Identifiers

PII
S1350448799001390;

Publishing Information

Journal Title
Radiation Measurements
Journal Volume
31
Journal Issue
1-6
Journal Page Range
p. 301-306
ISSN
1350-4487
CODEN
RMEAEP

Optional Information

Copyright
Copyright (c) 1999 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.