Absorbed dose response of Al2O3 dosimeter irradiated by 60Co γ spectrum source capture and collimators

  • 1. School of Physics Science and Engineering, Sun Yat-Sen Univ., Guangzhou (China)

Description

Objective: To investigated the absorbed dose response of Al2O3 dosimeter in water phantom irradiated by 60Co γ spectrum source. Methods: The EGSnrc simulation program code DOSRZnrc was used to calculate the absorbed dose of the Al2O3 dosimeter and that of the equivalent volume of water in the corresponding position, as well as the absorbed dose conversion factor, irradiated by 60Co photon beams in a water phantom. Simulations were done for a cylindrical geometry dosimeter (diameter 0.4 cm and height 0.1 cm) and the dosimeter was placed at the centre of the water phantom at different depths. Results: The average absorbed dose conversion factor is 1.143±0.006 and changes little with the depth of the dosimeter in the water phantom, and the deviation is less than 1.0%. Conclusion: The absorbed dose response of Al2O3 dosimeter irradiated by 60Co γ spectrum source is steady and is independent on the depth of the dosimeter in water phantom in this research. (authors)

Additional details

Publishing Information

Journal Title
Chinese Journal of Radiological Medicine and Protection
Journal Volume
29
Journal Issue
1
Journal Page Range
p. 91-93
ISSN
0254-5098

Optional Information

Notes
4 figs., 11 refs.