Calibration of ionization chamber for 18F and 68Ga

Description

In order to maintain the results of primary activity standardizations carried out in 2011 the LNMRI has determined the calibration factors for a pressurized 4π-ionization chamber for the nuclides 18F and 68Ga. This ionization chamber is coupled to a 6517A Keithley electrometer which is controlled by a homemade LabVIEW program. This paper will describe the main issues related to the calibration of an ionization chamber system for positron emitters and short half-life radionuclides such as timing, current measurement, background, decay, and 226Ra check source measurements. - Highlights: • Calibration figures were determined for 18F and 68Ge/68Ga. • The main components of uncertainty were determined. • Use long-lived 68Ge/68Ga as replacement for 18F in quality control of calibrators. • Sent four 68Ge/68Ga reference solutions to manufacturers of radionuclide calibrators. • A standardized 68Ge/68Ga solution was submitted to the SIR at BIPM

Availability note (English)

Available from http://dx.doi.org/10.1016/j.apradiso.2013.11.137

Additional details

Identifiers

DOI
10.1016/j.apradiso.2013.11.137;
PII
S0969-8043(13)00587-3;

Publishing Information

Journal Title
Applied Radiation and Isotopes
Journal Volume
87
Journal Page Range
p. 188-191
ISSN
0969-8043
CODEN
ARISEF

Conference

Title
19. international conference on radionuclide metrology and its applications
Dates
17-21 Jun 2013
Place
Antwerp (Belgium)

Optional Information

Copyright
Copyright (c) 2014 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.