Published January 2006 | Version v1
Journal article

QA of intensity-modulated beams using dynamic MLC log files

  • 1. Department of Radiation Oncology, Yashoda Cancer Institute, Secunderabad (India)

Description

To evaluate the utility of Dynalog file information for planar dose verification in IMRT QA, a program is developed to convert Dynalog file data to DMLC field files. For this study, five predefined fluencies are planned and delivered using Varian, Eclipse 3D planning system and 6MV photon beam of Varian, Clinac DMX linear accelerator. To measure planar dose distribution, Kodak, EDR2 films are exposed in similar setup as planning setup. Dynalog files are recorded for each delivery and converted into DMLC field files using in-house program. Delivered dose distributions are calculated using DMLC field files from Dynalog files. Planned, Measured and Delivered dose distributions are compared using gamma evaluation in Scanditronix, Omni Pro IMRT software. The Planned and Delivered planar dose distributions agree within 2% dose difference and 2 mm DTA. Measured dose distributions agree within 4% dose difference and 4 mm DTA with Planned dose distribution. Our results show Dynalog file as a promising tool for dynamic IMRT QA. (author)

Additional details

Publishing Information

Journal Title
Journal of Medical Physics
Journal Volume
31
Journal Issue
1
Journal Page Range
p. 36-41
CODEN
JMPHFE

INIS

Country of Publication
India
Country of Input or Organization
India
INIS RN
40016763
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
COLLIMATORS; GAMMA RADIATION; QUALITY ASSURANCE; RADIATION DOSE DISTRIBUTIONS; RADIOTHERAPY
Descriptors DEC
ELECTROMAGNETIC RADIATION; IONIZING RADIATIONS; MEDICINE; NUCLEAR MEDICINE; RADIATIONS; RADIOLOGY; THERAPY

Optional Information

Notes
7 refs., 9 figs., 1 tab.