Radiochromic film dosimetry: Reflection vs transmission scanning

  • 1. Medical Physics Laboratories of the Ioannina University School and the Ioannina University Hospital, 451.10 Ioannina (Greece)

Description

The objective of this study was to compare the transmission and the reflection scanning reading modes of radiochromic MD-55 films of a document scanner. The use of the red channel of the red-green-blue images results in measurement accuracy that did not differ between the two reading modes. On the other hand, the scanner uniformity and temporal stability, the long-term stability of the film readings and the dynamic range is superior using the reflection mode. The dynamic range of dose measurements can be extended up to least 500 Gy using the alternative color channels by both reading techniques. Similar backscattering dose factors close to high Z inhomogeneities were found using both scanning modes. In conclusion, the use of reflection scanning mode of MD55 films was superior to the traditionally used transmission mode

Additional details

Identifiers

Publishing Information

Journal Title
Medical Physics
Journal Volume
35
Journal Issue
6
Journal Page Range
p. 2308-2311
ISSN
0094-2405
CODEN
MPHYA6

INIS

Country of Publication
United States
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
40003271
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY;
Descriptors DEI
ACCURACY; BACKSCATTERING; DOSEMETERS; FILM DOSIMETRY; IMAGES; RADIATION DOSES; REFLECTION; STABILITY
Descriptors DEC
DOSES; DOSIMETRY; MEASURING INSTRUMENTS; SCATTERING

Optional Information

Notes
(c) 2008 American Association of Physicists in Medicine