Published March 2014 | Version v1
Journal article

The use of iterative reconstruction technique in low-dose CT angiography of renal artery

  • 1. Department of Radiology, First Affiliated Hospital, School of Medicine, Zhejiang University, Hangzhou (China)

Description

Objective: To assess the feasibility of computed tomography angiography (CTA) in renal artery using the iterative reconstruction technique(iDose4) in Brilliance iCT (Philips Ltd.). Methods: 90 patients (body weight 60-70 kg) supposed to undergo renal artery CTA were selected and randomly divided into three groups: A(120 kV, 120 mAs), B(120 kV, 60 mAs) and C(100 kV, 100 mAs). CTA images from Group A were obtained using filtered back projection and those from the rest groups were generated by iDose4. Then CT number of renal artery, noise (denoted as standard deviation SD), signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) of images for each group were measured and recorded. Images were post-processed using volume reformation (VR) 3D imaging and image quality were assessed in five-level grade (1-5). Results: The five parameters (CT number of renal artery, SD value, SNR, CNR and VR grade) of group A and group B were statistically insignificant (P > 0.05). While SD values of group C were comparable to those of group A and B, the rest four parameters of group C were apparently larger, which was statistically significant (tCT value = -12.326, tSNR = -7.136, tCNR = -7.649, ZVR = -2.690, PVR < 0.05). Conclusions: Image quality could still fit the diagnosis requirements using iDose4 when mAs decreased lower than 50%. The image quality increases when radiation dose is halved at 100 kVp kV. (authors)

Additional details

Publishing Information

Journal Title
Chinese Journal of Radiological Medicine and Protection
Journal Volume
34
Journal Issue
3
Journal Page Range
p. 228-230
ISSN
0254-5098

Optional Information

Notes
1 tab., 14 refs.