Published November 1985
| Version v1
Journal article
Direct-calibration method for determination of aluminum in serum is comparable with the protein-precipitation technique
Description
The authors compared results by two methods for serum Al determination: matrix modification with direct calibration in a stabilized-temperature platform furnace and a technique involving extraction with nitric acid before atomic absorption spectrometry. The two methods gave similar results with use of either a deuterium or a Zeeman system of background correction, but gave different slopes for standard additions, depending on the Al content of the serum, an effect not seen with aqueous solutions. These differences do not affect the accuracy of the Al determination up to 150 micrograms/L
Additional details
Publishing Information
- Journal Title
- Clin. Chem. (Winston-Salem, N.C.)
- Journal Issue
- no.11
- Series
- Clin. Chem. (Winston-Salem, N.C.).
- ISSN
- 0009-9147
- CODEN
- CLCHA
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 17043808
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Descriptors DEI
- ABSORPTION SPECTROSCOPY; ALUMINIUM COMPOUNDS; BLOOD CHEMISTRY; CALIBRATION; DEUTERIUM; MAN; PRECIPITATION
- Descriptors DEC
- ANIMALS; BIOCHEMISTRY; CHEMISTRY; HYDROGEN ISOTOPES; ISOTOPES; LIGHT NUCLEI; MAMMALS; NUCLEI; ODD-ODD NUCLEI; PRIMATES; SEPARATION PROCESSES; SPECTROSCOPY; STABLE ISOTOPES; VERTEBRATES