Published June 1, 1994
| Version v1
Journal article
Method for measuring the light output of scintillating glass shards
Creators
- 1. Pacific Northwest Laboratory, Richland, WA 99352 (United States)
Description
In order to determine the physical processes that govern the production of light in scintillating glasses, it is necessary to work with as-produced samples which have not been subjected to heat treatment, polishing, or other preparation. A combination of an optical system, which is insensitive to sample size and shape, and a Compton backscatter coincidence technique provides a method which gives precise and reproducible results when working with such samples. ((orig.))
Additional details
Publishing Information
- Journal Title
- Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
- Journal Volume
- 345
- Journal Issue
- 1
- Journal Page Range
- p. 95-98.
- ISSN
- 0168-9002
- CODEN
- NIMAER
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- Netherlands
- INIS RN
- 25074404
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- BACKSCATTERING; COINCIDENCE METHODS; COMPTON EFFECT; EFFICIENCY; GAMMA DETECTION; GLASS; KEV RANGE 100-1000; MEASURING METHODS; NAI DETECTORS; OPTICAL SYSTEMS; PULSES; SAMPLE HOLDERS; SCINTILLATIONS; SOLID SCINTILLATION DETECTORS; TESTING
- Descriptors DEC
- BASIC INTERACTIONS; COUNTING TECHNIQUES; DETECTION; ELASTIC SCATTERING; ELECTROMAGNETIC INTERACTIONS; ENERGY RANGE; INTERACTIONS; KEV RANGE; MEASURING INSTRUMENTS; RADIATION DETECTION; RADIATION DETECTORS; SCATTERING; SCINTILLATION COUNTERS