Neutron spectrum measurements in the aluminum oxide filtered beam facility at the Brookhaven Medical Research Reactor

  • 1. EG and G Idaho, Inc., Idaho Falls (USA)

Description

Neutron spectrum measurements were performed on the aluminum oxide filter installed in the Brookhaven Medical Research Reactor (BMRR). For these measurements, activation foils were irradiated at the exit port of the beam facility. A technique based on dominant resonances in selected activation reactions was used to measure the epithermal neutron spectrum. The fast and intermediate-energy ranges of the neutron spectrum were measured by threshold reactions and 10B-shielded 235U fission reactions. Neutron spectral data were derived from the activation data by two approaches: (1) a short analysis which yields neutron flux values at the energies of the dominant or primary resonances in the epithermal activation reactions and integral flux data for neutrons above corresponding threshold or pseudo-threshold energies, and (2) the longer analysis which utilized all the activation data in a full-spectrum, unfolding process using the FERRET spectrum adjustment code. This paper gives a brief description of the measurement techniques, analysis methods, and the results obtained

Additional details

Publishing Information

Publisher
Plenum Press.
Imprint Place
New York, NY (USA)
Imprint Title
Neutron beam design, development, and performance for neutron capture therapy
Imprint Pagination
339 p.
Journal Page Range
p. 235-245.

Conference

Title
Workshop on neutron beam design, development and performance for neutron capture therapy.
Dates
30-31 Mar 1989.
Place
Cambridge, MA (USA).

Optional Information

Secondary number(s)
CONF-8903150--.