Published April 2018 | Version v1
Journal article

Consistency of photon emission intensities for efficiency calibration of gamma-ray spectrometers in the energy range from 20 keV to 80 keV

  • 1. CEA, LIST, Laboratoire National Henri Becquerel (LNE-LNHB), Bât.602 PC111, CEA-Saclay, Gif-sur-Yvette Cedex, 91191 (France)

Description

Highlights: • Efficiency calibration of four HPGe detectors using standard radioactive sources. • Accurate processing of the full-energy peaks and Monte Carlo simulation. • Absolute photon emission intensities derived for Ba-133. • Absolute K X-ray emission measured for Sr-85, Y-88, Cd-109, Eu-152 and Ho-166 m. - Abstract: The efficiency calibration for different high-purity germanium detectors in the low-energy range was established by the conventional method, using standard radioactive sources. The peak shapes were carefully analysed taking account of natural linewidth, full-energy width at half maximum and scattering. Complementary information was obtained by Monte Carlo simulation using the PENELOPE code, after optimization of the geometrical parameters. This was used to measure photon emission intensities of some low-energy emitting radionuclides, including 133Ba, and compared to the tabulated values.

Availability note (English)

Available from http://dx.doi.org/10.1016/j.apradiso.2017.07.006

Additional details

Identifiers

DOI
10.1016/j.apradiso.2017.07.006;
PII
S0969804317303111;

Publishing Information

Journal Title
Applied Radiation and Isotopes
Journal Volume
134
Journal Page Range
p. 131-136
ISSN
0969-8043
CODEN
ARISEF

Optional Information

Notes
© 2017 Elsevier Ltd. All rights reserved.