Published February 1993
| Version v1
Journal article
Digital X-ray detector based on a CCD matrix
Description
The application of charge-coupled devices (CCD) as position-sensitive detectors for X-ray radiation is discussed. The experimental scheme contains a videotransmission camera to detect pulsed X-ray radiation and Vektor analog-to-digital converters and computer interface. It is shown that the sensitivity range of a K1200TsM1 matrix to bremsstrahlung with an average energy of 75 keV is 2.7·10-8-0.8·10-6 J/cm2. The spatial resolution of the resultant X-ray image is discussed. 5 refs., 5 figs
Additional details
Publishing Information
- Journal Title
- Instruments and Experimental Techniques
- Journal Volume
- 35
- Journal Issue
- 4
- Journal Page Range
- p. 689-692.
- ISSN
- 0020-4412
- CODEN
- INETAK
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 25011094
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Translation
- Descriptors DEI
- BREMSSTRAHLUNG; CHARGE-COUPLED DEVICES; POSITION SENSITIVE DETECTORS; PULSES; RANGE; SENSITIVITY; SPATIAL RESOLUTION; X-RAY DETECTION
- Descriptors DEC
- DETECTION; ELECTROMAGNETIC RADIATION; MEASURING INSTRUMENTS; RADIATION DETECTION; RADIATION DETECTORS; RADIATIONS; RESOLUTION; SEMICONDUCTOR DEVICES
Optional Information
- Notes
- Cover-to-cover Translation of Pribory I Tekhnika Eksperimenta (USSR); Translated from Pribory i Tekhnika Eksperimenta; No. 4, 165-168(Jul-Aug 1992).