Published November 2019
| Version v1
Journal article
Calibration of 3 MV Tandetron™ accelerator over nominal energy range
Creators
- 1. Institute for Physics and Power Engineering, Bondarenko sq. 1, Obninsk, 249033 (Russian Federation)
Description
The method based on the measurement of proton beam energies using the non-resonant direct-capture reaction 16O(p,γ)17F* with both charged and neutral projectiles was applied to perform the energy calibration of the 3 MV Tandetron™ accelerator throughout the nominal energy range. The uncertainty of the beam energy was found to be well within the accuracy specified by the manufacturer. No deviations from the linearity were revealed.
Availability note (English)
Available from http://dx.doi.org/10.1016/j.nimb.2019.08.031Additional details
Identifiers
- DOI
- 10.1016/j.nimb.2019.08.031;
- PII
- S0168583X19305920;
Publishing Information
- Journal Title
- Nuclear Instruments and Methods in Physics Research. Section B, Beam Interactions with Materials and Atoms
- Journal Volume
- 459
- Journal Page Range
- p. 81-83
- ISSN
- 0168-583X
- CODEN
- NIMBEU
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 54122728
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S43: PARTICLE ACCELERATORS;
- Descriptors DEI
- ACCELERATORS; CALIBRATION; OXYGEN 16; PROJECTILES; PROTON BEAMS; VOLTMETERS
- Descriptors DEC
- BEAMS; ELECTRIC MEASURING INSTRUMENTS; ELECTRICAL EQUIPMENT; EQUIPMENT; EVEN-EVEN NUCLEI; ISOTOPES; LIGHT NUCLEI; MEASURING INSTRUMENTS; NUCLEI; NUCLEON BEAMS; OXYGEN ISOTOPES; PARTICLE BEAMS; STABLE ISOTOPES
Optional Information