Feasibility of 320-row area detector CT coronary angiography using 40 mL of contrast material: assessment of image quality and diagnostic accuracy
- 1. Seoul National University Hospital, Department of Radiology, Seoul (Korea, Republic of)
Description
To assess the image quality and diagnostic accuracy of 320-row area detector CT (320-ADCT) coronary angiography using 40 mL of contrast material in comparison with 60-mL protocol. This retrospective study included 183 patients who underwent 320-ADCT coronary angiography using 40 mL of contrast and additional 183 sex- and body mass index-matched patients using 60 mL of contrast constituting the control group. Both groups used the same 5-mL/sec injection rate. Quantitative image quality measurements and diagnostic accuracies were calculated and compared. Mean attenuation and contrast-to-noise ratio (CNR) at the aorta and all coronary arteries were lower in the 40-mL group than in the 60-mL group (all, p < 0.05), except for the CNR at proximal coronary arteries at 100 kVp (p = 0.073). However, the proportion of coronary segments with vessel attenuation >250 HU was not different between groups (all, p > 0.05), except for distal coronary arteries at 80 kVp (p = 0.001). Furthermore, there were no differences in per-patient and per-segment diagnostic accuracies between the groups (all, p > 0.05). 320-ADCT coronary angiography using 40 mL of contrast showed image quality and diagnostic accuracy comparable to the 60-mL protocol, demonstrating the clinical feasibility of lowering the risk of contrast-induced nephropathy through contrast volume reduction. (orig.)
Availability note (English)
Available from: http://dx.doi.org/10.1007/s00330-016-4275-9Additional details
Identifiers
Publishing Information
- Journal Title
- European Radiology
- Journal Volume
- 26
- Journal Issue
- 11
- Journal Page Range
- p. 3802-3810
- ISSN
- 0938-7994
- CODEN
- EURAE3
INIS
- Country of Publication
- Germany
- Country of Input or Organization
- Germany
- INIS RN
- 48000303
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ACCURACY; AGE GROUPS; BLOOD CIRCULATION; BLOOD FLOW; COMPARATIVE EVALUATIONS; COMPUTERIZED TOMOGRAPHY; CONTRAST MEDIA; CORONARIES; DIAGNOSIS; FEASIBILITY STUDIES; PLAQUE FORMATION; SENSITIVITY; SIGNAL-TO-NOISE RATIO; SPECIFICITY; WEIGHT
- Descriptors DEC
- ARTERIES; BLOOD VESSELS; BODY; CARDIOVASCULAR SYSTEM; DIAGNOSTIC TECHNIQUES; DIMENSIONLESS NUMBERS; EVALUATION; ORGANS; TOMOGRAPHY