Published January 1993 | Version v1
Journal article

Angular dependent effect on solid state personal dosemeter

Description

Angular depended effect on solid state personal dosemeter. Angular dependent effect of solid state personal dosemeter mydose mini CdS (made by Aloka)and TLD UD-802P CaSO4 : Tm (made by Chiyoda) have been done. Response of both type of dosemeters toward radiation depends on angle of radiation direction. Highest responses of the dosemeters are achieved at angle of Oo. Their responses decrease by increasing angle. Nevertheless, for radiation protection purposes both dosemeters are suffiently good. Decrease is response toward radiation was only 4% for mydose mini CdS and 2% for TLD UD-802P CaSO4 : Tm when the range the angle exposure is between -300 to + 300. (author). 6 refs., 3 figs

Additional details

Additional titles

Original title (Indonesian)
Pengaruh sudut terhadap tanggapan dosimeter perorangan zat padat

Publishing Information

Journal Title
Fusi
Journal Volume
1
Journal Issue
1
Journal Page Range
p. 13-19.
ISSN
0853-9936
CODEN
FSIIED