The effect of radiation dose reduction on clinical image quality in chest radiography of premature neonates using a dual-side readout technique computed radiography system
- 1. Dept. of Medical Physics and Biomedical Engineering, Sahlgrenska Univ. Hospital, SE-413 45 Gothenburg (Sweden)
- 2. Dept. of Radiation Physics, Univ. of Gothenburg, SE-413 45 Gothenburg (Sweden)
- 3. Dept. of Biomedical Engineering, Varberg Hospital, SE-432 81 Varberg (Sweden)
- 4. Dept. of Paediatric Radiology, The Queen Silvia Children's Hospital, Sahlgrenska Univ. Hospital, SE-416 85 Gothenburg (Sweden)
Description
The purpose of the present study was to investigate if the exposure could be reduced from the clinical setting (resulting in an effective dose of 8 μSv for a neonate of weight 0.7 kg and height 25 cm at a tube voltage of 90 kV) without negatively influencing the image quality for a dual-side readout technique computed radiography (CR) system in chest radiography of premature neonates. Chest radiographs of premature neonates were acquired with the double-side readout technique CR system. The images underwent simulated dose reduction in steps of 20% to represent five different radiation dose levels. Four image quality criteria, related to the visibility of important anatomical structures, were used in a visual grading study where five experienced radiologists rated how well the criteria were fulfilled for all images. When reducing the radiation dose, a decrease in image quality could be observed already at the 80% dose level for all the structures. The results indicate that a decrease in exposure from the clinically used setting affects the image quality negatively for the CR system. (authors)
Availability note (English)
Available from doi: http://dx.doi.org/10.1093/rpd/ncq072Additional details
Identifiers
- DOI
- 10.1093/rpd/ncq072;
Publishing Information
- Journal Title
- Radiation Protection Dosimetry
- Journal Volume
- 139
- Journal Issue
- 1-3
- Journal Page Range
- p. 275-280
- ISSN
- 0144-8420
Conference
- Title
- 3. Malmoe Conference on Medical Imaging - Optimisation in X-ray and molecular imaging
- Dates
- 25-27 Jun 2009
- Place
- Malmoe (Sweden)
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- France
- INIS RN
- 43102241
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- BODY BURDEN; CHEST; COMPUTERIZED TOMOGRAPHY; HUMAN POPULATIONS; IMAGES; INFANTS; NEONATES; PARTURITION; RADIATION DOSES; RADIATION PROTECTION; READOUT SYSTEMS; SENSITIVITY; SIMULATION; SPECIFICITY
- Descriptors DEC
- AGE GROUPS; ANIMALS; BODY; CHILDREN; DIAGNOSTIC TECHNIQUES; DOSES; MAMMALS; MAN; POPULATIONS; PRIMATES; TOMOGRAPHY; VERTEBRATES
Optional Information
- Notes
- 19 refs