Published 2010 | Version v1
Journal article

The effect of radiation dose reduction on clinical image quality in chest radiography of premature neonates using a dual-side readout technique computed radiography system

  • 1. Dept. of Medical Physics and Biomedical Engineering, Sahlgrenska Univ. Hospital, SE-413 45 Gothenburg (Sweden)
  • 2. Dept. of Radiation Physics, Univ. of Gothenburg, SE-413 45 Gothenburg (Sweden)
  • 3. Dept. of Biomedical Engineering, Varberg Hospital, SE-432 81 Varberg (Sweden)
  • 4. Dept. of Paediatric Radiology, The Queen Silvia Children's Hospital, Sahlgrenska Univ. Hospital, SE-416 85 Gothenburg (Sweden)

Description

The purpose of the present study was to investigate if the exposure could be reduced from the clinical setting (resulting in an effective dose of 8 μSv for a neonate of weight 0.7 kg and height 25 cm at a tube voltage of 90 kV) without negatively influencing the image quality for a dual-side readout technique computed radiography (CR) system in chest radiography of premature neonates. Chest radiographs of premature neonates were acquired with the double-side readout technique CR system. The images underwent simulated dose reduction in steps of 20% to represent five different radiation dose levels. Four image quality criteria, related to the visibility of important anatomical structures, were used in a visual grading study where five experienced radiologists rated how well the criteria were fulfilled for all images. When reducing the radiation dose, a decrease in image quality could be observed already at the 80% dose level for all the structures. The results indicate that a decrease in exposure from the clinically used setting affects the image quality negatively for the CR system. (authors)

Availability note (English)

Available from doi: http://dx.doi.org/10.1093/rpd/ncq072

Additional details

Identifiers

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
139
Journal Issue
1-3
Journal Page Range
p. 275-280
ISSN
0144-8420

Conference

Title
3. Malmoe Conference on Medical Imaging - Optimisation in X-ray and molecular imaging
Dates
25-27 Jun 2009
Place
Malmoe (Sweden)

INIS

Country of Publication
United Kingdom
Country of Input or Organization
France
INIS RN
43102241
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Resource subtype / Literary indicator
Conference
Descriptors DEI
BODY BURDEN; CHEST; COMPUTERIZED TOMOGRAPHY; HUMAN POPULATIONS; IMAGES; INFANTS; NEONATES; PARTURITION; RADIATION DOSES; RADIATION PROTECTION; READOUT SYSTEMS; SENSITIVITY; SIMULATION; SPECIFICITY
Descriptors DEC
AGE GROUPS; ANIMALS; BODY; CHILDREN; DIAGNOSTIC TECHNIQUES; DOSES; MAMMALS; MAN; POPULATIONS; PRIMATES; TOMOGRAPHY; VERTEBRATES

Optional Information

Notes
19 refs