Published July 1975 | Version v1
Journal article

Atomization in graphite-furnace atomic absorption spectrometry. Peak-height method vs. integration method of measuring absorbance: carbon rod atomizer 63

  • 1. Carleton Univ., Ottawa

Description

Oscilloscopic traces of transient atomic absorption signals generated during continuous heating of a Carbon Rod Atomizer model 63 show features which are characteristic of the element being atomized. This research was undertaken to determine the significance and usefulness of the two analytically significant parameters, absorbance maximum and integrated absorbance. For measuring integrated absorbance, an electronic integrating control unit consisting of a timing circuit, a lock-in amplifier, and a digital voltmeter, which functions as a direct absorbance x second readout, has been designed, developed, and successfully tested. Oscilloscopic and recorder traces of the absorbance maximum and digital display of the integrated absorbance are simultaneously obtained. For the elements studied, Cd, Zn, Cu, Al, Sn, Mo, and V, the detection limits and the precision obtained are practically identical for both methods of measurements. The sensitivities by the integration method are about the same as, or less than, those obtained by the peak-height method, whereas the calibration curves by the former are generally linear over wider ranges of concentrations. (U.S.)

Additional details

Identifiers

Publishing Information

Journal Title
Analytical Chemistry
Journal Volume
47
Journal Issue
8
Series
Anal. Chem.
Journal Page Range
1240-1249
ISSN
0003-2700

Optional Information

Notes
Updated automatically by Metadata and Full-Text Enrichment Agent