Published 1985 | Version v1
Report Open

Time properties of Ge(Li) detectors of multidetector correlation set up

Description

Time resolution of semiconductor Ge(Li) detectors of true-coaxial type and selection of trigger forming current parameters are investigated. Optimization of the time charrels permits one to obtain symmetric form of the prompt time curve with 2 taU 0 < or approximately 11 ns width for each pair of detectors. The time curves of Ge(Li) detectors have hees measured in γ-γ-coincidence mode using the 60Co source. The results obtained permit to use these detectors in multidetector correlation spectrometer to study angular correlations and to determine half-life of isomeric states in nanosecond region (Tsub(1/2)>3 ns)

Availability note (English)

MF available from INIS under the Report Number.

Files

16078313.pdf

Files (177.1 kB)

Name Size Download all
md5:7253dd2cfec4d0cc99a3ba90f77aa8b7
177.1 kB Preview Download

System files (19.3 kB)

Name Size Download all

Additional details

Additional titles

Original title (Russian)
Временные свойства Ге(Ли)-детекторов многодетекторной корреляционной установки

Publishing Information

Imprint Pagination
5 p.
Report number
JINR-R--6-85-83

Optional Information

Notes
1 ref.; 3 figs.; 2 tabs.