Published November 1, 2019 | Version v1
Journal article

Characterization of dynamic collimation mechanisms for helical CT scans with direct measurements

  • 1. Department of Radiology, Division of Diagnostic Imaging Physics, Massachusetts General Hospital, 55 Fruit Street, Boston, MA 02114 (United States)

Description

Dynamic collimation is an important dose reduction mechanism for helical CT scans, especially for modern wide-beam scanner models. Its implementation and efficacy need to be studied to optimize CT scan protocols and to reduce unnecessary patient dose.

The purpose of this study is to evaluate dynamic beam collimation for modern wide-beam CT scanners with direct measurements and to estimate the efficacy for dose reduction.

By using a linear-array solid state detector, primary x-ray beam coverage was measured for four CT scanner models: GE Revolution CT, Siemens Somatom Force, Philips iQon, and GE LightSpeed VCT. Supported independently from patient table, the detector remained stationary at the isocenter during helical scans. Data lines were recorded every 0.24 ms throughout one entire helical scan, with a spatial resolution of 0.8 mm along the craniocaudal direction. The measurements were repeated for various scan parameters related to dynamic collimation, including beam collimation width, pitch, rotation time, and scan length. The recorded beam coverage area was used as a surrogate to total primary dose, to model different dynamic collimation mechanisms. The directly measured total radiation range was compared to table travel distance and nominal scan length which equals to the ratio between DLP and CTDIvol. Equations to calculate the percentage dose reduction with dynamic collimation were derived for different mechanisms.

Three different dynamic collimation mechanisms were revealed and related linear model parameters were reported for different helical scan parameters. The nominal scan length used to calculate DLP was shown to vary for different dynamic collimation mechanisms. For typical head and abdomen scans with nominal scan lengths of 17.5 cm and 25 cm, percentage dose reduction from dynamic collimation ranged from 2% to 32%.

In conclusion, with direct measurements of primary x-ray beam coverage, dynamic collimation mechanisms and related dose reduction effects were characterized for four modern wide-beam CT scanners. (paper)

Availability note (English)

Available from http://dx.doi.org/10.1088/1361-6560/ab3eaa

Additional details

Identifiers

Publishing Information

Journal Title
Physics in Medicine and Biology
Journal Volume
64
Journal Issue
21
Journal Page Range
[9 p.]
ISSN
0031-9155
CODEN
PHMBA7

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
52004097
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
BEAM DYNAMICS; GERMANIUM; NUCLEAR INDUSTRY; RADIATION DOSES; SPATIAL RESOLUTION; X RADIATION
Descriptors DEC
DOSES; DYNAMICS; ELECTROMAGNETIC RADIATION; ELEMENTS; INDUSTRY; IONIZING RADIATIONS; MECHANICS; METALS; RADIATIONS; RESOLUTION