Published June 2016 | Version v1
Journal article

eHXI: a permanently installed, hard x-ray imager for the National Ignition Facility

  • 1. Lawrence Livermore National Laboratory, 7000 East Ave., Livermore, California 94550 (United States)
  • 2. Laboratory for Laser Energetics, University of Rochester, 250 East River Rd., Rochester, New York 14623 (United States)

Description

We have designed and built a multi-pinhole imaging system for high energy x-rays (≥ 50 keV) that is permanently installed in the equatorial plane outside of the target chamber at the National Ignition Facility (NIF). It records absolutely-calibrated, time-integrated x-ray images with the same line-of-sight as the multi-channel, spatially integrating hard x-ray detector FFLEX [McDonald et al., Rev. Sci. Instrum. 75 (2004) 3753], having a side view of indirect-drive inertial confinement fusion (ICF) implosion targets. The equatorial hard x-ray imager (eHXI) has recorded images on the majority of ICF implosion experiments since May 2011. eHXI provides valuable information on hot electron distribution in hohlraum experiments, target alignment, potential hohlraum drive asymmetries and serves as a long term reference for the FFLEX diagnostics.

Availability note (English)

Available from http://dx.doi.org/10.1088/1748-0221/11/06/P06010

Additional details

Publishing Information

Journal Title
Journal of Instrumentation
Journal Volume
11
Journal Issue
06
Journal Page Range
p. P06010
ISSN
1748-0221