Efficiency calibration of a mini-orange type β-spectrometer by the β--spectrum of 90Sr

Description

A specific method for efficiency calibration of a mini-orange type β-spectrometer by means of the continuous β--spectrum of 90Sr and the conversion electron spectrum of 207Bi in the energy range from 500 to 2200 keV has been elaborated. In the experiment typical SmCo5 magnets (6A and 8A) were used. An accuracy of efficiency determination was 5-10 %. (author)

Availability note (English)

Available from INIS in electronic form

Files

34060986.pdf

Files (433.8 kB)

Name Size Download all
md5:690d893f329c03038f951da7e8b8f3e6
433.8 kB Preview Download

System files (21.6 kB)

Name Size Download all

Additional details

Additional titles

Original title (Russian)
Калибровка β-спектрометра типа 'мини-апельсин' по эффективности с помощью β

Publishing Information

Imprint Pagination
12 p.
Report number
JINR-R--6-2002-159

Optional Information

Notes
5 refs., 2 figs.